Table 2 Detailed results of the compression tests with the cube-shaped PLA specimen. Mean value (XÌ…) and standard deviation (SÌ…)

From: Biomechanical validation of additively manufactured artificial femoral bones

    

E (MPa)

σy (MPa)

σp (MPa)

Infill

Test direction

n

XÌ…

SÌ…

XÌ…

SÌ…

XÌ…

SÌ…

Printed cubes

30%

X/Y

5

61.3

17.0

2.3

0.9

3.5

0.7

Z

5

179.3

28.1

4.3

0.7

6.4

0.7

40%

X/Y

5

187.8

24.6

5.7

0.5

7.9

0.7

Z

5

392.2

82.7

6.9

0.9

10.6

1.4

50%

X/Y

5

385.2

66.4

8.2

1.0

12.1

1.3

Z

5

1084.1

237.4

11.2

3.3

17.7

1.8