Skip to main content

Table 4 Maximum values of subsidence and maximum insertion force (Fmax), as well as surface strain (εmax) for the respective specimens during the biomechanical tests (a testing was aborted due to suspected fracture)

From: Biomechanical validation of additively manufactured artificial femoral bones

 

Subsidence (mm)

εmax (%)

Fmax (kN)

Human Femur

HB1

35

0.019

3.9

HB2

30

0.164

8.5

HB3

25

–

4.2a

Composite Bones

PC

35

0.131

3.8

PLA

35

0.144

4.7

ABS

40

0.052

4.7

PMMA

25

–

5.0