Table 1 The table shows the average values and their standard deviation for FWHM, FWTM, width at − 100 HU, and AE in CT number for the different beam energies. p-values represent comparing metric scores to the 120 kV case. Artifact regions are shown in Fig. 2

From: Needle artifact reduction during interventional CT procedures using a silver filter

Metric

100 kV

120 kV

135 kV

120 kV- Ag

Trajectory artifact

Needle tip artifact

Needle artifact

Trajectory artifact

Needle tip artifact

Needle artifact

Trajectory artifact

Needle tip artifact

Needle artifact

Trajectory artifact

Needle tip artifact

Needle artifact

FWHM (mm)

Difference % with 120 kV

4.91 ± 0.15

(p = 0.026)

11.3

2.37 ± 0.07 (p = 0.702)

−2.9

2.17 ± 0.04 (p = 0.885)

− 0.9

4.41 ± 0.09

2.44 ± 0.07

2.19 ± 0.02

4.51 ± 0.09

(p = 0.892)

2.3

2.31 ± 0.03

(p = 0.310)

− 5.3

1.98 ± 0.03

(p = 0.003)

− 9.6

4.25 ± 0.12 (p = 0.677)

− 3.6

2.08 ± 0.02 (p = 0.007)

− 14.8

1.88 ± 0.04 (p < 0.001)

− 14.2

FWTM (mm)

Difference % with 120 kV

10.06 ± 0.59

(p = 0.500)

− 10.8

3.81 ± 0.13 (p = 0.679)

− 5.9

7.42 ± 0.73 (p = 0.937)

− 3.5

11.28 ±

1.24

4.05 ± 0.34

7.69 ±

0.17

9.47 ± 0.29 (p = 0.296)

− 16.0

3.53 ± 0.02 (p = 0.225)

− 12.8

5.36 ± 0.34 (p = 0.016)

− 30.3

9.03 ± 0.41 (p = 0.169)

− 19.9

3.11 ± 0.08 (p = 0.029)

− 23.2

3.59 ± 0.01 (p < 0.001)

− 53.3

Width at −100 HU (mm)

Difference % with 120 kV

7.85 ± 0.30 (p = 0.569)

4.8

n/a

n/a

7.49 ± 0.23

n/a

n/a

5.97 ± 0.35 (p = 0.010)

− 20.3

n/a

n/a

4.89 ± 0.03 (p < 0.001)

− 34.7

n/a

n/a

AE in CT number (HU)

Difference % with 120 kV

479 ± 10

(p = 0.887)

2.4

1768 ± 69

(p = 0.130)

11.4

4379 ± 209

(p = 0.091)

10.5

468 ± 8

1587 ± 42

3963 ± 42

311 ± 29 (p < 0.001)

− 33.5

1113 ± 93 (p = 0.003)

− 29.9

4363 ± 145

(p = 0.150)

10.1

240 ± 3

(p < 0.001)

− 48.7

737 ± 48 (p < 0.001)

− 53.6

4361 ± 4 (p = 0.153)

10.0

  1. The negative symbol in the percentage difference indicates metal artifacts reduction with the reference beam of 120 kV